一氧化碳?xì)怏w檢測(cè)管1LL Carbon monoxide    的詳細(xì)介紹一氧化碳?xì)怏w檢測(cè)管1LL Carbon monoxide
Carbonmonoxide                  No.1LL 
MeasuringRange
5 to 50 ppm
Number of PumpStrokes
2
CorrectionFactor
1
SamplingTime
2 minutes per pumpstroke
DetectingLimit
1 ppm (n=2)
ColorChange
Yellow → Blackishbrown
ReactionPrinciple
CO +Na2Pd(SO3)2  → Pd + CO2  + SO2  + Na2SO3
Coefficient ofVariation15% (for 5 to 10ppm), 10% (for 10 to 50 ppm)ShelfLife3 YearsCorrectionsforSubstance
Concentration
Interference
Change color by itself
Acetylene
1/50 time
Plus error
Produces dark brownstain
Sulfurdioxide
1/10 time
Plus error
Produces dark brownstain
Nitrogendioxide
40 ppm
Plus error
Produces redstain
Hydrogensulfide
1/5 time
Plus error
Produces dark brownstain